frequencies, we can see that the percentage of the droplets below 5.79 mm was higher and the median diameter lower for the 2.4-MHz frequency,although ventilation was higher.

    The increase in vibration frequency decreased the size of emitted droplets. This is difficult to observe with the 1.63-MHz frequency compared to 1.5MHz because ventilation was at its highest.

   Table 5 presents the influence of a coupling liquid on solution temperature after nebulization for 10 min. With nebulizers B and C, the increase in temperature was moderate because of the presence of a coupling liquid that prevented contact between the solution and the transducer. With nebulizer A,the temperature of the solution after nebulization for 10 min was higher than 40 .The coupling liquid absorbed the energy produced by ultrasonic vibrations and cancelled out any heating of the solution, which often occurs with ultrasonic nebulizers

(11–13). The rise in temperature might be responsible for drug degradation.For example,after ultrasonic nebulization,Ip et al. (13) reported denaturing of a protein (recombinant methionyl interferon consensus) that could be avoided by not

heating the solution. In the same way, Cipolla et al.(12) noted an aggregation of rH DNase because of heightened temperature.

   The use of ultrasonic nebulizers at high vibration intensity and low ventilation gives satisfactory nebulization of solutions as regards the size of emitted droplets, the quantity of drug nebulized, and the nebulization time. A low ventilation level is preferable,on the one hand, for the emission of the

smaller-size droplets required to administer the drug deep into the lungs and, on the other hand, for better compatibility with patient administration.High intensity is preferred because it increases the quantity of drug nebulized and therefore the quantity of drug likely to reach the lungs. The presence

of a coupling liquid is interesting because it prevents the increase in temperature that could degrade some drugs. High vibration frequency is better because it makes it possible to obtain smaller-size droplets.

    For a given ultrasonic nebulizer, the vibration frequency of the transducer is fixed. Only vibration intensity and ventilation level have to be defined to nebulize a given solution with an ultrasonic nebulizer.In contrast, with jet nebulization, a nebulizer with given geometric characteristics has to be chosen: liquid and air tube orifice diameters, size and shape of the reservoir, shape and position of the impaction system. All these have to be associated with defined airflow and pressure conditions.

CONCLUSION

An ultrasonic nebulizer is efficient, with high vibration frequency, for drug administration into the lungs. High vibration intensity associated with a low ventilation level is preferable for the delivery of drugs deep into the lungs. Using a coupling liquid that transmits ultrasonic vibrations to the solution and limits any increase in temperature is favorable. This is particularly interesting for thermosensitive substances that could be used with this type of nebulizer.

To maintain a constant quantity of drug for the patient at each administration, it is necessary to standardize operating conditions: vibration intensity and ventilation level therefore have to be defined.

It is indispensable that solutions for nebulization should be subjected to thorough pharmaceutical trials before use to define the best-adapted nebulizer(s) and the operating conditions for those retained. In the interest of public health, when a Marketing Authorization Application file is requested for a liquid preparation to be nebulized and its droplet size greatly influences drug therapeutic activity, it seems important to us not only to define the formulation, but also to associate with it the proper nebulizer(s) and administration conditions.This is the case for both specialties Pentacarinat* (Bellon, Neuilly/Seine, France) containing pentamidine and Pulmozyme* (Roche, Neuilly/Seine,France) containing dornase alpha, which obtained the Marketing Authorization Application file in 1989 and 1994.

上一篇:搅拌釜反应器英文文献和中文翻译
下一篇:在线机器测量系统英文文献和中文翻译

AngularJS技术介绍英文文献和中文翻译

开关电源水冷却系统英文文献和中文翻译

减数分裂和基因重组英文文献和中文翻译

太阳能最大功率点追踪和...

移动码头的泊位分配问题英文文献和中文翻译

虚拟船舶装配集成建模方...

中学生科学探究中对等论...

提高教育质量,构建大學生...

上海居民的社会参与研究

从政策角度谈黑龙江對俄...

压疮高危人群的标准化中...

基于Joomla平台的计算机学院网站设计与开发

STC89C52单片机NRF24L01的无线病房呼叫系统设计

酵母菌发酵生产天然香料...

浅论职工思想政治工作茬...

AES算法GPU协处理下分组加...

浅谈高校行政管理人员的...